TECHNICAL SPECIFICATIONS

DIMENSIONS

 
Size: 13.5" x 12" x 7.5"
34.29 cm x 30.48 cm x 19.05 cm
Weight: 28.5 lbs
13.0 Kg
Portability: I.V pole mountable, free standing or handle on top of unit.
   
POWER  
Power: 115-120V or 230V AC input
Battery Type: Rechargeable lead acid
Fluid Pump: Roller type peristaltic pump 10 -1000 ml/min in 10 ml/min steps
Heater: Maximum applied power, 1400 watts
Control Panel Display: 5" X 2.5" (12.7 cm X 6.35 cm) Electroluminescent type with water proof touch pad
 
Made in USA. Hyperthermia Pump has been evaluated with respect to electrical safety, fire and mechanical hazards only in accordance with UL2601-1, CAN/CSA C22.2 No. 601.1- M90, and EN 60601-1: 1990 + A1: 1993 + A11: 1993 + A12: 1993 + A2: 1995 + A13: 1996 Electrical Compliance:EN 60601-1, UL 2601-1, CSA/C22.2 - No. 601.1-M90
 
Disclaimer: The Belmont™ Hyperthermia Pump is a heater/circulator which provides a safe and effective method for delivering fluid warmed to an operator controlled temperature and flow rate. The Pump is to be used by or under the supervision of a physician who determines the type of infusate fluid, its concentration, temperature and flow rate, and the duration of the infusion/warming procedure. The physician also determines the appropriateness of a particular hyperthermic treatment in treating disease.