TECHNICAL SPECIFICATIONS

DIMENSIONS

 
Size  
Power Module:

4.0” x 4.0” x 13.0” 
10.16 cm x 10.16 cm x 33.02 cm

Heater Unit: 1.5” x 5.2” x 1.5”  
3.81 cm x 13.21 cm x 3.81 cm
Weight  
Power Module: 12 lbs
5.45 kg
Heater Unit: 3.5 oz. (7 oz., with cable) 
99.22 g (198.45 g., with cable)
       
POWER  
AC Input Voltage/Frequency: 100-120V~; 50-60 Hz
 
Maximum Power: 200VA
Electrical Compliance, fire and mechanical hazards safety: UL60601-1, CAN/CSA C22.2-ND601.1-M90
   
BATTERY PACK  
Nominal Voltage/Capacity: 12V, 15AH
Battery Charging: Automatic when AC power is connected, with or without heating.
Recharge time: 8 hrs.
Heating Capacity: Warms at least 5 liters of fluid to 38°C from 20°C input in one battery cycle
 
   
ENVIRONMENTAL  
Operating Temperature, Humidity: 10° to 35°C, 10% to 70% RH
Shock and Vibration: ISTA Test procedure 1, 1996
Electromagnetic Compatibility: EN60601-1-2
Protection from water Ingress: IPX1, Drip Proof
 
   
OPERATING PARAMETERS  
Flow Rate – AC  
Input Temperature 20°C: Max 100 ml/min
Input Temperature 10°C: Max 50 ml/min
Flow Rate – Battery Operation  
Input Temperature 20°C: Max 65 ml/min
Input Temperature 10°C: Max 35 ml/min
 
Output Temperature: 38° +/-3°C
Prime Volume: 7.5 ml with valves and tubing
   
ALARMS/ALERTS  
Low Flow  Low Temperature
Check Connection, Press Power On Empty Set/Check for Air
Low Battery/Recharge Now    No Heat/Check Connection
Over Current  Over Temperature
Probe Fault   System Reset
   
DISPOSABLE SET  
General: Non-Pyrogenic Fluid Path, Single-Patient Use, Latex-free, DEHP-free
Pressure Regulating Valve in Inflow Tubing: Prevents pressure> 400mg Hg from reaching the set or the patient.
Check Valve in Outflow Tubing: Prevents back flow from the patient
Venting Membrane: Microporous membrane vents air from crystalloid solutions
Anti-entrainment Valve built into the Set: Prevents air entrainment into the set.